FDAJune 8, 2023device

HF Resection Electrodes - HF resection electrodes are part of a resectoscope system for endoscopic diagnosis and treatment in urological or gynecological applications. 64 Models: 1) A22201C 2) A22201D 3) A22202C 4) A22203C 5) A22205C 6) A22205D 7) A22206C 8) A22207C 9) A22211C 10) A22221C 11) A22...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip

What to do

FDA enforcement status: Ongoing

Brands named

olympus

UPCs

140427610366481404276103665514042761036662140427610366861404276103670914042761036716140427610367231404276103674714042761036761140427610294661404276103043114042761029770

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
HF Resection Electrodes - HF resection electrodes are part of a resectoscope system for endoscopic diagnosis and treatment in urological or gynecological applications. 64 Models: 1) A22201C 2) A22201D 3) A22202C 4) A22203C 5) A22205C 6) A22205D 7) A22206C 8) A22207C 9) A22211C 10) A22221C 11) A22... — Recall Details · AllClear