FDAJuly 22, 2024device

SlideView VS200 slide viewer,

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

What to do

FDA enforcement status: Ongoing

Brands named

evident scientificevident

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.