FDAJuly 14, 2025device

Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

What to do

FDA enforcement status: Ongoing

Brands named

maquet

UPCs

040467681451770404676814518404046768145191040467681452070404676814521404046768145221

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.