WATCHMAN TruSeal Access System DBL, OUS, Material Number (UPN) M635TS70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOrca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.2026-05-12
- FDAIntera Oncology INTERA 3000 Hepatic Artery Infusion Pump, AP-0300H, implantable drug delivery device2026-05-11
- FDABoston Scientific Model SC-4318, Clik X Anchor REF: M365SC431802026-04-29
- FDABoston Scientific CSK Electrodes: REF: CSK-TC10, CSK Electrode, STAINLESS STEEL, 10 cm x 28 ga (0.36 mm) REF: CSK-TC10-3M,CSK Electrode, STAINLESS STEEL, 3M Cable, 10 cm x 28 ga (0.36 mm); REF: CSK-TC15, CSK Electrode, STAINLESS STEEL, 15 cm x 28 ga (0.36 mm); REF: CSK-TC15-3M, CSK Electrode, STA...2026-04-29
- FDABoston Scientific TCN Electrodes: REF: TCN-5, TCN Electrode, NITINOL, 5 cm x 0.45 mm (0.0175 in); REF:TCN-10, TCN Electrode, NITINOL, 10 cm x 0.45 mm (0.0175 in) REF:TCN-15, TCN Electrode, NITINOL, 15 cm x 0.45 mm (0.0175 in); REF:TCN-20, TCN Electrode, NITINOL, 20 cm x 0.45 mm (0.0175 in) REF: T...2026-04-29
- FDALUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.52026-03-30
- FDALUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.52026-03-30
- FDABoston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.2026-03-19
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →