FDAJanuary 4, 2018device

BioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV) is an automated test for use on the VIDAS¿ system to detect abnormal operation of the VIDAS and mini VIDAS instrument pipette mechanisms and optical systems. Specifically the m...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS¿ and mini VIDAS¿ system due to a decrease of pipetted volume.

What to do

FDA enforcement status: Terminated

Brands named

biomerieux

UPCs

0357302613357303573026133580035730261092570357302611406003573026114084035730261404270357302602967803573026351663035730261216310357302603029203573026553319

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
BioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV) is an automated test for use on the VIDAS¿ system to detect abnormal operation of the VIDAS and mini VIDAS instrument pipette mechanisms and optical systems. Specifically the m... — Recall Details · AllClear