FDAJuly 29, 2025device

WATCHMAN TruSeal Access System SGL, US, Material Number (UPN) M635TU70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

What to do

FDA enforcement status: Ongoing

Brands named

boston scientificboston

UPCs

08714729965701

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.