Dimension Vista¿ CTNI Sample Diluent, KD692, SMN# 10445205

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.

What to do

FDA enforcement status: Terminated

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare