FDAMay 15, 2020device

(1) CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink Data Management System software version 16.0.3 Siemens Material Number (SMN): 11314337 Product Usage: are clinical laboratory information modules which consolidate test orders and ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

What to do

FDA enforcement status: Terminated

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare