FDAMay 15, 2020device
(1) CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink Data Management System software version 16.0.3 Siemens Material Number (SMN): 11314337 Product Usage: are clinical laboratory information modules which consolidate test orders and ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review
What to do
FDA enforcement status: Terminated
Brands named
siemens healthcare diagnosticssiemenssiemens healthcare
UPCs
0063041422063500630414237268
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAChlorthalidone Tablets, USP, 25 mg, [100 or 1000] Tablets pr bottle, Rx only, Manufactured by: Inventia Healthcare Limited, Additional Ambernath, M.I.D.C., Ambernath (East) - 421506, INDIA. Distributed by: Risiong Pharma Holdings, Inc., East Brunswick, NJ 08816. NDC 100-tablet bottle: 64980-599-0...2026-06-05
- FDASiemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay2026-05-07
- CPSCBayer Recalls 6 mL Size Afrin Original Nasal Spray Bottles Due to Risk of Serious Injury or Illness from Child Poisoning; Violates Mandatory Standard for Child-Resistant Packaging2026-04-30
- FDAdirectCHECK ACT-LR whole blood quality control, level 2 Model/Catalog Number: DCGLR-22026-04-30
- FDABrand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.2026-04-22
- FDABioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.2026-04-22
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;2026-04-02
- FDABiofire Spotfire Respiratory/Sore Throat Panel REF: 4234852026-04-01