FDAMay 15, 2020device

(1) Atellica Data Manager software version 1.0 Siemens Material Number (SMN): 11314237 (2) Atellica Data Manager software version 1.1 Siemens Material Number (SMN): 11316888 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test ord...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

What to do

FDA enforcement status: Terminated

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare