FDAAugust 26, 2019device
Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.
What to do
FDA enforcement status: Terminated
Brands named
livanova deutschlandlivanova
UPCs
04033817900993
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABrand Name: SenTiva DUO" Product Name: VNS Therapy¿ SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-D Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external prog...2025-01-03
- FDABrand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming syst...2025-01-03
- FDASenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).2024-12-12
- FDASenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).2024-12-12
- FDAEssenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software2024-10-18
- FDA3T Heater Cooler System, REF: 16-02-852023-08-03
- FDASorin Centrifugal Pump System, CP5, Drive Unit (Catalogue Item #60-01-04, UDI: 04033817900948) used in an HLM S5/C5 System configuration. A cardiopulmonary bypass speed control device.2021-12-17
- FDAVagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model 303-30 PereniaDURA Lead (3.0 mm electrode inner diameter)2021-12-02