FDAJune 27, 2019device

AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.

What to do

FDA enforcement status: Terminated

Brands named

angiodynamics