FDAJune 24, 2024device

Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR EU-DIV/4096600-01, PURITAN BENNETT 560 VENTILATOR JAPAN/4096600-02, PURITAN BENNETT 560 VENTILATOR APAC/4096600-03, Puritan Bennett 560 Ventilator LATAM/4096600-04, P...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway

What to do

FDA enforcement status: Ongoing