FDAJuly 12, 2023device
The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASONIC GASTROVIDEOSCOPE, ULTRASONIC COLONOVIDEOSCOPE: GF-UC140P-AL5, GF-UCT140-AL5, GF-UE160-AL5, GF-UCT180, GF-UM20, GF-UM130, GF-UMQ130, GF-UM160, GF-UC160P-OL5, GF-UCT160-OL5
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body fluid could backflow into the air/water channel of the ultrasonic endoscopes during the procedure.
What to do
FDA enforcement status: Ongoing
Brands named
olympus
UPCs
04953170355929
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21122026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2118.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2110. May be included with the Olympus OER-ELITE.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21112026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116.2026-03-25
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →