FDAMay 19, 2026device

TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).

What to do

FDA enforcement status: Ongoing

Brands named

howmedica osteonicshowmedica

UPCs

07613327026061