FDAJune 14, 2024device

Optima MR360, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.

What to do

FDA enforcement status: Ongoing

Brands named

ge medical systems

UPCs

082427040158082427040159082427040131082427040134082427040136082427040138082427220034082427040156082427120128082427200082082427010067082427240051