FDAAugust 8, 2025device

RayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStation 11A Service Pack 2, Software Version: 11.0.3.116. 4. RayStation 11A Service Pack 3, Software Version: 11.0.4.15. Produc...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

What to do

FDA enforcement status: Ongoing

Brands named

raysearch laboratories abraysearchraysearch laboratories

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
RayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStation 11A Service Pack 2, Software Version: 11.0.3.116. 4. RayStation 11A Service Pack 3, Software Version: 11.0.4.15. Produc... — Recall Details · AllClear