FDAAugust 8, 2025device

RayStation with the following product descriptions: 1. RayStation 6 Service Pack 1, Software Version: 6.1.1.2. 2. RayStation 6 Service Pack 2, Software Version: 6.2.0.7. 3. RayStation 6 Service Pack 3, Software Version: 6.3.0.6. Product Description: Radiation Therapy Treatment Planning System.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

What to do

FDA enforcement status: Ongoing