FDAJuly 29, 2015device

KimVent* Wet Pak* Closed Suction System for Adults, 14 F, T-Piece (Product Code 22056); KimVent* Wet Pak* Closed Suction System for Adults, 16 F, T-Piece (Product Code 22556); KimVent* Wet Pak* Closed Suction System for Adults, 14 F, T-Piece (Product Code 22012356) A tracheobronchial suction cath...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

What to do

FDA enforcement status: Terminated

Brands named

halyard healthhalyard

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
KimVent* Wet Pak* Closed Suction System for Adults, 14 F, T-Piece (Product Code 22056); KimVent* Wet Pak* Closed Suction System for Adults, 16 F, T-Piece (Product Code 22556); KimVent* Wet Pak* Closed Suction System for Adults, 14 F, T-Piece (Product Code 22012356) A tracheobronchial suction cath... — Recall Details · AllClear