FDASeptember 23, 2019device

TOMTEC-ARENA TTA2;software version no. 2.20 and lower; Picture archiving and communications system. Software Versions and applicable UDI: TTA 1.2 release prior to UDI requirement TTA2.00.01 release prior to UDI requirement TTA2.11.01 release prior to UDI requirement TTA2.11.02 release prior to UD...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The firm discovered a software issue associated with the Image-Com module/clinical application package (CAP) interface for 3D application in the ARENA software, where measurements from one study might be stored to another study.

What to do

FDA enforcement status: Ongoing

Brands named

tomtec imaging systemstomtectomtec imaging