FDAAugust 1, 2025device
Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Mo...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
What to do
FDA enforcement status: Ongoing
Brands named
philips medical systems nederland b vphilipsphilips medical
UPCs
008848380590300088483805904700884838059054008848380590610088483805908500884838054189008848380541960088483805420200884838054219008848380542400088483805422600884838059115
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAPhilips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.2026-06-05
- FDAGSMS Incorporated, NIACIN EXTENDED-RELEASE TABLETS, USP, 1,000 MG, 90 tablets, Rx only, Manufactured by Kremers Urban Pharmaceuticals Inc., a subsidiary of Lannett, Inc., Seymour, IN 47274, Packaged by GSMS Incorporated, Camarillo, CA 93012. NDC 51407-268-90.2026-05-29
- FDAPhilips Avalon Fetal Monitor, FM 30 Part numberM2703A2026-05-26
- FDAPhilips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A2026-05-26
- FDAPhilips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # ...2026-05-18
- FDAPhilips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Elev...2026-05-18
- CPSCGenerac Power Systems Expands Recall of Generac and DR Power Electric Start Pressure Washers Due to Risk of Serious Injury or Death from Carbon Monoxide Hazard; Includes Additional Models2026-05-14
- FDANicotine Transdermal System Patch, 14 mg, 14 patches per box, Aveva, Distributed by: RUGBY LABORATORIES, Indianapolis, IN 46268, www.majorpharmaceuticals.com, NDC 0536-5895-88.2026-05-14
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →