FDAMay 8, 2020device

uCT 760 System, Computed Tomography X-ray System

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

What to do

FDA enforcement status: Terminated

Brands named

shanghai united imaging healthcareshanghaishanghai united