FDAJuly 20, 2015device
Neptune 2 Waste Management System Docking Station (120V and 230V), New and Refurbished Product Usage: The Stryker Docking Station (Docker) is a component of the Neptune 2 Waste Management System. The Stryker Rover, another component of the system, is a mobile unit used to suction and collect flui...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The Fluid Coupling Sleeve, a component within the Neptune Waste Management System Docker, may be out of specification. This can cut or dislodge the PTFE seal and cause a leak between the Rover and Docker couplers during docking or leaking from the Rover coupler after docking. The potential hazard is waste fluid and/or fresh water dispensed from the Neptune 2 Rover at the incorrect time.
What to do
FDA enforcement status: Terminated
Brands named
stryker instruments divstrykerstryker instruments
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDANeptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System2026-05-06
- FDADaig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic Tip Steerable, Product Number 6DYNTP001;2026-04-10
- FDABARD Dynamic Tip Steerable, Product Number 200344; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDADaig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic XT Steerable, Product Number 201104; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →