FDAJuly 12, 2023device
nanoDot D2DNS, for custom calibrate, OTO, Model Number 03033-OTO; radiation monitoring dosimeter used with the microSTAR readers
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
What to do
FDA enforcement status: Completed
Brands named
landauer
UPCs
0860003399903
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAmicroStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110, and 201), 10016-000 (marketed in EMEA as 101), 10036-000, 10044-000, 10055-000, BC30058, and BC300592024-02-08
- FDALandauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-0002023-11-15
- FDAQC (Cs-137) screened nanoDot D2DNS, Model Number 03067-OTO; radiation monitoring dosimeter used with the microSTAR readers2023-07-12
- FDAnanoDot D2DNN, Model Numbers: a) 18105-000; b) 18105-KIT; radiation monitoring dosimeter used with the microSTAR readers2023-07-12
- FDACalibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 18120-KIT; c) 18120-SET; radiation monitoring dosimeter used with the microSTAR readers2023-07-12
- FDAReader, nanoDot adapter, EU Model Number VINLADA003; radiation monitoring dosimeter used with the microSTAR readers2023-07-12
- FDAnanoDot D2DNS, Model Numbers: a) 03055-1MO; b) 03055-3MO; c) 03055-OTO; d) 03055-SMO; radiation monitoring dosimeter used with the microSTAR readers2023-07-12
- FDAQC (Cs-137) nanoDot D2DNS, Model Numbers: a) 04225-000 (EMEA number VINLNAN005); b) 04225-KIT; c) 04225-SET; radiation monitoring dosimeter used with the microSTAR readers2023-07-12