FDAJune 4, 2018device
Leica Microsystems products labeled as the following: Leica M320 Microscope Systems with Article Numbers 10448419, 10448420, 10448485, 10448486 Product Usage: Intended use of this device is an optical instrument for magnifying and illuminating specimens. It can be applied for observation and docu...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Possible failure of screws which hold the gas spring fixation bracket in place. Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight of the optics carrier.
What to do
FDA enforcement status: Terminated
Brands named
leica microsystemsleica
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAHistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.2025-04-02
- FDABrand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.012025-01-30
- FDALeica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c) 14901850001R0501, d) 1491850UVUL; for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM1950, Material Numbers: a) 14047742467, b) 1491950C1US, c) 1491950C2US, d) 1491950C3US; for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c) 14903050S02, d) 14903050S03; for freezing and sectioning tissue samples2024-09-04