BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320 ): " OCTAVIUS Detector 1600 SRS (T10056) (UDI-DI: EPTWT100560) " OCTAVIUS Detector 1600 MR (T10057) (UDI-...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.
What to do
FDA enforcement status: Ongoing
Brands named
ptw freiburgptwptw freiburg