FDAJuly 28, 2016device

VITROS Immunodiagnostic Products TSH Reagent Pack, REF/Catalog Number/Product Code 191 2997, Unique Device Identifier No. 10758750000227, IVD --- Ortho-Clinical Diagnostics Pencoed Bridgend CF35 5PZ UK For the in vitro quantitative measurement of thyroid stimulating hormone (TSH) in human serum a...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A field correction was issued by the firm for VITROS TSH reagent lot 5040 due to a higher than expected frequency of customer calibration curves falling outside of the calibration quality parameters and therefore unusable to process samples. Customers using VITROS TSH Lot 5040 will potentially experience an unexpected increase in the unsuccessful calibrations potentially causing a delay in testing and reporting results for VITROS TSH.

What to do

FDA enforcement status: Terminated

Brands named

ortho clinical diagnosticsorthoortho clinical

UPCs

10758750000227

Recall history