FDAJuly 13, 2023device

ACUSON Redwood 2.0 ultrasound systems, REF: 11503314

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.

What to do

FDA enforcement status: Ongoing