FDAMarch 1, 2018device
VIDAS LH, Reference Numbers 30406 and 30406-01. Poduct Usage: VIDAS LH is an automated quantitative test for use on the VIDAS family instruments for the determination of human luteinizing hormone in human serum or plasma (lithium heparin), using the ELFA technique (Enzyme Linked Fluorescent Assay)
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specific lots.
What to do
FDA enforcement status: Terminated
Brands named
biomerieux
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAVITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 ...2026-01-05
- FDAVITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only2025-06-26
- FDAVITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST...2024-03-14
- FDAMYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects microbiology laboratory instruments for ID/AST and blood culture activities and the laboratory information system (LIS).2023-02-06
- FDAVITEK 2 REAGENT GN TEST KIT VTK2 20 CARDS, CATALOG 213412022-11-09
- FDAVITEK 2 REAGENT GP TEST KIT VTK2 20 CARDS, CATALOG 213422022-11-09
- FDATEMPO REAGENTS TEMPO CONTROL KIT , CATALOG 800002022-11-09
- FDAVIDAS CLINICALVIDAS VARICEL. ZOSTER IGG 60 T, CATALOG 302172022-11-09
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →