FDAAugust 13, 2025device

Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model Number: 728324.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

What to do

FDA enforcement status: Ongoing