FDAAugust 17, 2023device

Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 86...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients

What to do

FDA enforcement status: Ongoing

Brands named

philips north americaphilipsphilips north

UPCs

0088483810459400884838112070

Recall history