Gallant HF Cardiac Resynchronization Therapy Defibrillator: HF CRT-D REF CDHFA500Q Neutrino HF Cardiac Resynchronization Therapy Defibrillator: NxT HF CRT-D REF CDHFA600Q Entrant HF Cardiac Resynchronization Therapy Defibrillator : HF CRT-D REF CDHFA300Q The primary function of ICDs and CRT-Ds ar...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

What to do

FDA enforcement status: Ongoing

Brands named

abbott medicalabbott

UPCs

05415067032010

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Gallant HF Cardiac Resynchronization Therapy Defibrillator: HF CRT-D REF CDHFA500Q Neutrino HF Cardiac Resynchronization Therapy Defibrillator: NxT HF CRT-D REF CDHFA600Q Entrant HF Cardiac Resynchronization Therapy Defibrillator : HF CRT-D REF CDHFA300Q The primary function of ICDs and CRT-Ds ar... — Recall Details · AllClear