(1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica MAGNETOM Spectra Siemens Model Number:11060845 Used with the following MAGNETOM MRI Systems: MAGENTOM Amira 10836777 & 10838610; MAGENTOM Free.Max 11408424 & 11408425; MAGNETO...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used

What to do

FDA enforcement status: Ongoing

Brands named

siemens medical solutionssiemenssiemens medical

UPCs

0405686928174204056869006819040568690204880405686928175904056869281766040568698177304056869044521

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
(1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica MAGNETOM Spectra Siemens Model Number:11060845 Used with the following MAGNETOM MRI Systems: MAGENTOM Amira 10836777 & 10838610; MAGENTOM Free.Max 11408424 & 11408425; MAGNETO... — Recall Details · AllClear