FDAJuly 20, 2016device

Ingenuity Core128, Model number 728323 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

What to do

FDA enforcement status: Terminated

Brands named

philips medical systems clevelandphilipsphilips medical

UPCs

320004320053

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.