Philips DuraDiagnost stationary X-ray system

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels [21 CFR 1010.2., 21 CFR 1010.3(a), and 21 CFR 1020.30(3)].

What to do

FDA enforcement status: Terminated

Brands named

philips electronics north americaphilipsphilips electronics