FDAJune 30, 2020device

IntelliVue MX800 patient monitor, model no. 865240

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.

What to do

FDA enforcement status: Terminated

Brands named

philips north americaphilipsphilips north