FDAJune 30, 2020device
Alaris System PC Unit Model 8015, a modular infusion pump and monitoring system
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.
What to do
FDA enforcement status: Terminated
Brands named
carefusion 303carefusion
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABD ChloraPrep Clear, (2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA) 1 mL Applicator, 60 Applicators, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-31.2026-05-28
- FDABD ChloraPrep FREPP Clear,(2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA)) 1.5 mL Applicator, 20 Applicators, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-30.2026-05-28
- FDABD ChloraPrep FREPP Clear, (2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA) 1.5 mL Applicator, 20 Applicators, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-30.2026-05-28
- FDABD ChloraPrep Clear (2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA)), 1 mL Applicator, 60 Applicators per inner Carton, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-31.2026-05-28
- FDABD Pyxis MedStation 4000 REF: 303 Medication cabinet2026-04-30
- FDABD Pyxis MedStation ES REF: 323 Medication cabinet2026-04-30
- FDABD Pyxis Mini Main REF: 349 Medication cabinet2026-04-30
- FDABD Pyxis MedBank Tower REF: 1145-00 Medication cabinet2026-04-30
Own something like this?