FDAJune 6, 2017device

Draeger Medical Systems Infinity Acute Care System (M540) Catalog Numbers: MK31501/MK31701/MK31722

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Cockpits with revision index 06 or higher that contain 4GB RAM modules may not annunciate audio or visual alarms on the Cockpit and Central Station.

What to do

FDA enforcement status: Terminated

Brands named

draegar medical systemsdraegardraegar medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Draeger Medical Systems Infinity Acute Care System (M540) Catalog Numbers: MK31501/MK31701/MK31722 — Recall Details · AllClear