FDAJanuary 8, 2020device

SIEMENS Healthineers Luminos dRF Max Model: 10762471- Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical