FDAMay 15, 2015device

MedTest DX Control Reconstitution Fluid Kit configuration: 20 x 5 ml. Bottle: 5 ml amber glass Cap: 18 mm white seal plug seal rubber HC-DIL (Human Control Diluent) is reagent grade water filled into 5ml vials used to reconstitute Human controls HNC200 and HEC200. in vitro diagnostic.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Vial to vial variation in the fill volume that could affect control recovery issues and cause a control result to fall outside the suggested recovery ranges. An under-fill of the vials was found ranging as low as 4.63 ml. When used to reconstitute HNC/HEC, control values were shown to be out of range running on the high end.

What to do

FDA enforcement status: Terminated

Brands named

medtestdx

Recall history

No related federal recalls on record for this brand yet.