Fluorocell PLT reagent, Catalog #CD994563 Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
There is a potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%) results performed using the recalled lots may exhibit lower than expected PLT-F and IPF values. The issue may be recognized by a large discrepancy between the Impedance platelet (PLT-I) value and a false low PLT-F value. In most cases, the falsely decreased PLT-F results displayed a "PLT Abn Scattergram" message with asterisks beside the results indicating the data is unreliable. In rare instances, platelet flags may be absent. The PLT-I counts are unaffected. The PLT-F is a reflex test used to confirm PLT-I results due to flagging or low values. As such, the likely impact to patient results is low. Affected results display a discrepancy between the PLT-I and PLT-F, with lower than expected PLT-F results that do not match the patient's clinical picture.
What to do
FDA enforcement status: Terminated
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- CPSCGreenworks Tools Recalls 24V and 48V Kobalt Yard Power Tools with USB-C Batteries Due to Risk of Serious Injury from Fire Hazard2026-07-09
- FDAPhilips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.2026-06-05
- FDAPhilips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A2026-05-26
- FDAPhilips Avalon Fetal Monitor, FM 30 Part numberM2703A2026-05-26
- CPSCPetzl America Recalls SIMBA and SWAN EASYFIT Harnesses Due to Risk of Serious Injury or Death from Fall Hazard2026-05-14
- CPSCPetzl America Recalls ASTRO BOD FAST, ASTRO SIT FAST and CANYON GUIDE Harnesses Due to Risk of Serious Injury or Death from Fall Hazard2026-05-14
- FDADigitalDiagnost C90 Flex/Value/Chest/ER. Model Number: 712035.2026-05-01
- CPSCAcer America Corporation Recalls Two-Wheeled Folding Electric Due to Risk of Injury from Fall Hazard2026-04-30
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →