The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The Leica ASP200S/ASP300S Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.

What to do

FDA enforcement status: Terminated

Brands named

leica microsystemsleica