ACUSON SC2000 Ultrasound System with software versions VB10B and lower; Model 10433816; Product Usage: The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, A...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The ACUSON SC2000 ultrasound system considers uppercase/lowercase differences in the same patient name as unique patient instances when registered on the same ultrasound system. If these differences are not corrected at the time of registration, the system does not capture images or clips.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical