FDAMay 29, 2015device

Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images

What to do

FDA enforcement status: Terminated

Brands named

raysearch laboratories abraysearchraysearch laboratories