FDAJune 18, 2020device

Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 - Product Usage: This battery is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.

What to do

FDA enforcement status: Terminated

Brands named

philips north americaphilipsphilips north

UPCs

989803191341

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.