FDAMarch 26, 2020device

Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and trea...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.

What to do

FDA enforcement status: Terminated

Brands named

shanghai united imaging healthcareshanghaishanghai united