FDAJuly 20, 2020device

NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A second surgical procedure; 2) Implant replacement; 3) If placed in posterior lower jaw, the implant may damage the inferior alveolar nerve canal or perforate a small artery requiring immediate implant removal.

What to do

FDA enforcement status: Terminated

Brands named

nobel biocare services agnobelnobel biocare

UPCs

07332747073584

Recall history