FDAMay 17, 2018device

The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.

What to do

FDA enforcement status: Ongoing

Brands named

intuitive surgicalintuitive