Hitachi Scenaria CT system Product Usage: The SCENARIA CT system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table failed to stop when the control was released and contacted the CT gantry. The problem could not be duplicated at the site, so to avoid further issues while a root cause investigation was pursued, Hitachi choose to add an optional touch sensor to the gantry to add a redundant level of protection. It is conceivable that a patient s extremity may be caught between the tabletop and the gantry surface during uncontrolled motion or that the table contact of the gantry might cause internal damage to the unit.