FDAJune 29, 2016device

Broselow/Hinkle, PEDIATRIC EMERGENCY SYSTEM, Broselow Complete Set of Colored Pouches, Product Code 7730MOD

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The Oral Airway pouch within the PINK/RED Oxygen Delivery Module, 7700RAW, was incorrectly packaged with the 8 cm Berman Oral Airway. The pouch contained within the module is correctly labeled as 5 cm but actually contains an 8 cm Berman Oral Airway component. Therefore, the component within the pouch is incorrect. The PINK/RED Oxygen Delivery Module, 7700RAW, is intended to be used within the PINK/RED Pouch of the BROSELOW pediatric safety system. Three (3) lots (lot# 914873, lot# 914876 and lot #914878) of the PINK/RED Oxygen Delivery Module, 7700RAW, are affected by this voluntary Field Safety Corrective Action.

What to do

FDA enforcement status: Terminated