FDAJuly 14, 2020device

Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical